TECHNICAL SPECIFICATIONS UviLine 9400 UV-Visible
Wavelength range • 190-1100 nm
Light Source • Xenon
Bandpass • 4 nm
Incremental WL step • Reading: 0.1 nm • Setting : 1nm
Wavelenght accuracy • ± 1 nm
Wavelenght repeatability • ± 0,2 nm
Absorbance range • ± 3,300
Absorbance resolution • 0,001 Abs ou 0,1% T
Photometric accuracy • ± 0,003 Abs (0,5 Abs)
• ± 0,005 Abs (1,0 Abs)
• ± 0,010 Abs (2,0 Abs)
Stray light • ≤ 0.1 % to 220 nm (Nal)
340 nm (GG375)
400 nm (GG408)
Flatness baseline • ± 0,010 Abs
Scanning speed • Low – medium – fast
Update • Via USB port
Interface • 1 USB-A, 1 USB-B, 1 RS 232C
IP standard • IP 30 with drain in the cell compartment
Power supply • 110-220 V 50/60 Hz
specific country cable
Température (°C) • Use: 10 to 35 °C
/ Humidity • Storage –25 °C to 65 °C
Dimensions (L x W x H) • 404 x197 x 314 mm
Weight (net) • 4 Kg
ORDERING INFORMATION
• 70VI0500 : UviLine 9100 - VISIBLE 4 nm spectrophotometer
• 70VI0510 : UviLine 9400 - UV/VISIBLE 4nm spectrophotometer
• 80ZZ0035 : Pre-aligned spare lamp for UviLine 9100
• 70VI0600 : Automated 5+1 cell changer / instrument driven *
• 70VI0605 : 20/50/100 mm cell holder *
• 70VI0601 : Sipper / instrument driven *
• 70VI0607 : 16 mm tube holder. Quick-lock system
• 70VI0603 : 10mm cell holder thermostated (Peltier system) / instrument driven *
• 70VI0604 : 10mm cell holder (delivered standard with all UviLine series)
• 70VI0609 : 5mm-100mm universal tank holder & 16mm tube. Quick-lock system
• 70VI0610 : Sipper with Peltier effect
• 70VI0670 : PC LabPower software *